Category
Tiger Jazz
LED Driver
Online Service
Contact Us
|
MPQ-200
1~4 Output Medical Type 200W Medical MPQ-200 series
Features: * Universal AC input / Full range * Low leakage current <18 A * Protections: Short circuit / Overload / Over voltage / Over temperature Free air convection for 140W and forced air convection for 200W UL60601-1 medical safety approved * With power good and fail signal output * Built-in remote ON-OFF control * Built-in remote sense function * Fixed switching frequency at 100KHz * 3 years warranty Specifications:
Protections: SC / OL / OV · Universal AC input / Full range · PF>0.95@230VAC · PG / PF signal output · Protections: SC / OL / OV / OT · Leakage current ≤ 180µA · Built-in remote sense and remote ON/OFF control · Approvals: UL / CUL / TUV / CB / CE · 25 CFM forced air cooling · 100% full load burn-in test · 3 years warranty · AC input voltage range: 90 ~ 264VAC; 127 ~ 370VDC · AC Inrush Current: Cold start, 35A @ 115VAC, 60A @ 230VAC · DC Adjustment Range: CH1: -5 ~ +10% rated output voltage · Overload protection: 110% ~ 160% hiccup mode, auto-recovery · Over voltage protection: CH1: 115% ~ 140% rated output voltage · Leakage Current: Less than 150µA @ 264VAC · Setup, Rise, Hold Up Time: 1000ms, 20ms, 16ms @ full load and 230VAC · Withstand Voltage: I/P~O/P: 4KVAC, I/P-FG: 1.5KVAC, 1 minute · Working Temperature: -20 ~ 70°C (refer to output derating curve) · Safety standards: UL2601.1, TUV EN60601-1, IEC601-1 approved, design refer to UL 60950-1, TUV EN 60950-1 · EMC Standards: EN55011/EN55022 class B, EN61000-4-2,3,4,5,6,8,11, EN61000-3-2,3, EN60601-1-2, ENV50204
Related Product
|