Hangzhou Showland Power Co.,Ltd

RPS-60

RPS-60,
RPS-60
  • Product:
    RPS-60

  • Model Number:


  • Description:
     


    Learn more about products detailed description, pricing and service, Please send mail to us!

    sales@showlandpower.com
60W Single Output Medical Type RPS-60 series

Features:
*  4"x2"  miniature size
Universal AC input/Full range
Low leakage current<200uA
Protections: Short circuit / Overload / Over voltage
Cooling by free air convection
UL60601-1/IEC60601-1/EN60601-1 medical safety approved
UL60950-1/IEC60950-1/EN60950-1 ITE safety approved
No load power consumption<0.75W Fixed switch frequency at 100KHz
3 years warranty

Specifications:

MODEL

RPS-60-3.3

RPS-60-5

RPS-60-12

RPS-60-15

RPS-60-24  

RPS-60-48

OUTPUT

DC VOLTAGE

3.3V

5V

12V

15V

24V

48V

RATED CURRENT

10A

10A

5A

4A

2.5A

1.25A

CURRENT RANGE

0 ~ 11A 

0 ~ 11A

0 ~ 5.5A

0 ~ 4.4A

0 ~ 2.75A  

0 ~ 1.375A

RATED POWER

33W

50W

60W

60W

60W

60W

PEAK LOAD 10sec

36.3W

55W

66W

66W

66W

66W

RIPPLE & NOISE 

80mVp-p   

80mVp-p

120mVp-p

150mVp-p

240mVp-p  

300mVp-p

VOLTAGE ADJ. RANGE

3.1 ~ 3.6V 

4.75 ~ 5.5V   

11.4 ~ 13.2V  

13.5 ~ 16.5V  

22.8 ~ 27.6V

45.6 ~ 52.8V

VOLTAGE TOLERANCE

2.0%

2.0%

2.0%

2.0%

1.0%

1.0%

LINE REGULATION

0.5%

0.5%

0.5%

0.5%

0.5%

0.5%

LOAD REGULATION

1.0%

1.0%

1.0%

1.0%

1.0%

1.0%

SETUP, RISE TIME  

500ms, 30ms/230VAC    500ms, 30ms/115VAC at full load

HOLD UP TIME

50ms/230VAC    13ms/115VAC at full load

 

 

 

 

INPUT

VOLTAGE RANGE

90 ~ 264VAC   127 ~ 370VDC

FREQUENCY

47 ~ 63Hz

EFFICIENCY 

74%

79%

83%

84%

85%

86%

AC CURRENT 

1.8A/115VAC    1 A/230VAC

INRUSH CURRENT

COLD START 60A/230VAC    30A/115VAC

LEAKAGE CURRENT

For earth <200uA / 264VAC, For patient <100uA/264VAC

 

PROTECTION

OVER LOAD

115 ~ 150% rated output power

Protection type : Hiccup mode, recovers automatically after fault condition is removed

OVER VOLTAGE

3.8 ~ 5V   

5.75 ~ 6.75V  

13.8 ~ 16.2V

17.25 ~ 20.25V

28.4 ~ 32.4V

55.2 ~ 64.8V

Protection type : Shut down o/p voltage, re-power on to recover

 

 

 

ENVIRONMENT

WORKING TEMP.

-20 ~ +70oC (Refer to output load derating curve)

WORKING HUMIDITY

20 ~ 90% RH non-condensing

STORAGE TEMP., HUMIDITY

-40 ~ +85oC , 10 ~ 95% RH

TEMP. COEFFICIENT

0.03%/ oC (0 ~ 45oC ) 

VIBRATION

10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

 

 

 

 

 

SAFETY &EMC

SAFETY STANDARDS

UL60950-1, TUV EN60950-1, UL60601-1, TUV EN60601-1, IEC60601-1 approved

WITHSTAND VOLTAGE  

I/P-O/P:4KVAC I/P-FG:1.5KVAC O/P-FG:1.5KVAC

ISOLATION RESISTANCE

I/P-O/P, I/P-FG, O/P-FG:100M Ohms/500VDC

EMI CONDUCTION & RADIATION

Compliance to EN55011 (CISPR11), EN55022 (CISPR22) Class B

HARMONIC CURRENT

Compliance to EN61000-3-2,-3

EMS 

IMMUNITY

Compliance to EN61000-4-2,3,4,5,6,8,11; ENV50204, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A

MTBF

353.6Khrs min. MIL-HDBK-217F (25 )

DIMENSION

101.6*50.8*29mm (L*W*H)

PACKING  

0.15Kg; 96pcs/15.4Kg/0.89CUFT


 